| Model Number |
H7493895915200 |
| Medical Device Problem Code |
Kinked (1339)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/10/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device evaluated by mfr.: returned product consisted of an apex balloon catheter.The balloon was loosely folded.The shaft, hypotube, and bonds were microscopically and visually examined.The shaft was completely separated 119.4cm from the hub.The fractured/separated ends of the shaft were stretched and jagged which indicates the shaft separation was due to tensile forces.There is tip damage.There was no evidence of any material or manufacturing deficiencies contributing to the damage.There were numerous hypotube and shaft kinks.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2017.It was reported that shaft kink occurred.The target lesion was located in the right coronary artery.A 2.00mm x 15mm apex¿ balloon catheter was advanced but it was noted that the device delivery shaft was kinked.The procedure was completed with another 2.00mm x 15mm apex¿ balloon catheter.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed a broken shaft.
|
| |
|