• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CG4+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CG4+ CARTRIDGE Back to Search Results
Catalog Number 03P85-50
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Abdominal Pain (1685)
Date of Event 02/23/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling was evaluated during the investigation as pertaining to the event.The investigation was completed on 06/15/2017.Retain product was tested and functioning according to specification.Return product is not available for investigation.Investigation: the device history record (dhr) for this lot was reviewed.The lot passed finished goods (fg) release criteria.Forty retained cartridges from the lot were tested using whole blood (wb) and i-stat level 1 control (l1).Testing met the acceptance criteria found in q04.01.003 rev.X, appendix 1 - product complaint level 2 and level 3 investigation procedure.No deficiency has been identified.Assessment: the complaint investigation concluded there was no product deficiency and that the i-stat cartridge lot is meeting specification.The treatment for elevated lactate would be fluids, antibiotics, oxygen and observation.No treatment was administered based on the > 20 mmol/l result.
 
Event or Problem Description
On (b)(6) 2016, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges that yielded suspected discrepant lactate result on a (b)(6) year old male patient with epigastric abdominal pain.There was no additional patient information available at the time of this report.Return product is not available for investigation.(b)(6).At the time of the event there was no indication of a product malfunction based on the information provided and assessed by apoc, however this mdr is a retrospective filing in response to an observation from an fda inspection conducted may 8th to 12th, 2017 at abbott point of care (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CG4+ CARTRIDGE
Common Device Name
CG4+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key7141642
Report Number2245578-2017-00649
Device Sequence Number939184
Product Code KHP
UDI-Device Identifier10054749002266
UDI-Public10054749002266
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K982071
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2016
Device Catalogue Number03P85-50
Device Lot NumberD16007
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/22/2017
Initial Report FDA Received Date12/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-