| Catalog Number |
70-6529-R |
| Medical Device Problem Codes |
Electrical Shorting (2926); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/27/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analyzer was evaluated upon return to manufacturer.At magellan analyzer powered on successfully using ac adapter.Intermittent issues in powering analyzer on using batteries.Analyzer made crackling noises when powered on with batteries in the left battery chamber.No indication of injury or harm to the user.Late filing is part of (b)(4).
|
| |
|
Event or Problem Description
|
|
Instrument reported to generate 102 error codes.In addition, the ac adapter for analyzer became extremely hot.No user injury reported.
|
| |
|