Model Number N/A |
Device Problem
Insufficient Information (3190)
|
Patient Problems
Pulmonary Embolism (1498); Perforation of Vessels (2135); Anxiety (2328); No Information (3190); No Code Available (3191)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).The event is currently under investigation.A supplemental report will be submitted upon completion.
|
|
Event Description
|
It is alleged that the plaintiff received a gunther filter on (b)(6) 2009.It is alleged that the plaintiff was injured without further explanation.Hospital and medical records have been requested but not yet provided.
|
|
Manufacturer Narrative
|
It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿gunther tulip filter implanted".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
(b)(4).The event is currently under investigation.A supplemental report will be provided upon conclusion.
|
|
Event Description
|
This additional information received on 31/jan/2017 as follows: patient received an implant on (b)(6) 2009 via the right common femoral vein due to pulmonary embolism and upcoming orthopedic surgery.Patient is alleging mesenteric perforation, anxiety, mental anguish and stress due to the device.
|
|
Manufacturer Narrative
|
Additional information: investigation: investigation is reopened due to additional information provided.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'vena cava and mesenteric perforation,anxiety, mental anguish and stress-'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported anxiety, mental anguish and stress is directly related to the filter and unable to identify corresponding failure mode(s) at this time.(b)(4) devices in lot.No relevant notes on work order.No other complaints on lot.Product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|
Manufacturer Narrative
|
Product problem to adverse event and product problem.Malfunction to serious injury.Blank fields on this form indicate the information is unknown or unavailable, or unchanged.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating ¿vena cava and mesenteric perforation, anxiety, mental anguish and stress".Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Vena cava wall perforation is a known potential complication of vena cava filters.Both symptomatic and asymptomatic events have been reported.Among other causes, vena cava wall perforation may inadvertently be initiated by improper deployment, excessive force or manipulations near an implanted filter (e.G., a surgical procedure in the vicinity of a filter) and (or) procedures that involve other devices being passed through an in situ filter.There is a current debate in the published scientific literature on a differentiation between ivc wall perforation with and without clinical sequelae.E.G.Filter legs may be outside the contrast lumen on imaging without actually perforating the ivc wall (known as tenting) and with no clinical sequelae.In contrast, perforation of adjacent organs is reported with clinical sequelae.Unknown if the reported anxiety, mental anguish, and stress are directly related to the filter and unable to identify a corresponding failure mode at this point in time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
|
|