| Catalog Number |
02.08.IF1.06.RMLL |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/24/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reason of this deformation was attributed to some high viscosity cement debris remained on the tibial tray surface at the clipping interface with the poly.Batch review performed on (b)(6) 2017.Lot 168938: 15 items manufactured and released on (b)(6) 2017.Expiration date: (b)(6) 2022.No anomalies found related to the problem.To date, 1 item of the same lot have been already sold.
|
| |
|
Event or Problem Description
|
|
During the surgery the surgeon was not able to clip the inlay in the tibial tray.The surgeon removed the insert and found some cement debris on the tibial tray.The 8mm liner was deformed and for this a new liner 9mm was opened and implanted successfully.
|
| |
|