• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC APPLICATION INSTRUMENT FOR STERNAL ZIPFIX; CERCLAGE FIXATION Back to Search Results
Model Number 03.501.080
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that two (2) zipfix guns were slipping each time it was tensioned.The surgeon confirmed that this was not noticed during any specific procedure.He has noticed it over time with each continued use.He was still able to use the same device to complete a procedure.Procedure was completed with no negative impact to patient.No other information provided at this time.Concomitant devices reported: zipfix implant (part # unknown, lot # unknown, quantity unknown).This report is for one (1) application instrument for sternal zipfix.This is report 2 of 2 for complaint (b)(4).
 
Manufacturer Narrative
The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
A device history record review was performed: part no.: 03.501.080, lot no.: 7806881.Manufacturing location: (b)(4), release to warehouse date: 09.Mar.2012.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Service history review states: no service history review can be performed as part number 03.501.080 with lot number(s) 7806881 is a lot/batch controlled item.The service history review is unconfirmed.The customer reported the zipfix gun was slipping during tensioning.The repair technician reported the retaining nut was loose, and the trigger was sticking.Sticky trigger is the reason for repair.The cause of the issue is unknown.No parts were replaced.The item was repaired per the inspection sheet, passed synthes final inspection on 23-jan-2018 and will be returned to the customer upon completion of the service and repair process.Service record router completed through operation 30.Finalized service record will be archived in docusphere document management system.The evaluation was confirmed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APPLICATION INSTRUMENT FOR STERNAL ZIPFIX
Type of Device
CERCLAGE FIXATION
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key7143163
MDR Text Key95797984
Report Number2939274-2017-50406
Device Sequence Number1
Product Code JDQ
UDI-Device Identifier10887587010847
UDI-Public(01)10887587010847(10)7806881
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 11/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.501.080
Device Catalogue Number03.501.080
Device Lot Number7806881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2017
Initial Date FDA Received12/22/2017
Supplement Dates Manufacturer Received12/29/2017
01/18/2018
01/25/2018
Supplement Dates FDA Received01/17/2018
01/23/2018
01/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-