Model Number 1435-
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/01/2015
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product event summary: it was reported that the controller dc adapter did not make a smooth easy snap on connection.The controller dc adapter ((b)(4)) was returned for evaluation.However, the device associated with this product event exceeded the retention period as per current procedures and is therefore not available for analysis.A review of documentation records confirmed that the associated device met all requirements for release.Applicable risk documentation and experience with events similar circumstances were considered.A possible root cause of the inability of the controller dc adapter to make a smooth connection may include but not limited to a connector assembly failure and/or due to the handling of the device.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.
 
Event or Problem Description
It was reported that the controller dc adapter did not make a smooth easy snap on connection.The controller dc adapter was replaced.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7143478
Report Number3007042319-2017-05420
Device Sequence Number1300896
Product Code DSQ
UDI-Device Identifier00888707000109
UDI-Public00888707000109
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1435-
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2015
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2017
Initial Report FDA Received Date12/22/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-