| Model Number |
1435- |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/01/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: it was reported that the controller dc adapter did not make a smooth easy snap on connection.The controller dc adapter ((b)(4)) was returned for evaluation.However, the device associated with this product event exceeded the retention period as per current procedures and is therefore not available for analysis.A review of documentation records confirmed that the associated device met all requirements for release.Applicable risk documentation and experience with events similar circumstances were considered.A possible root cause of the inability of the controller dc adapter to make a smooth connection may include but not limited to a connector assembly failure and/or due to the handling of the device.This event was assessed and is being reported as part of a retrospective review of events, which was in response to an update to the mdr decision criteria.This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.
|
| |
|
Event or Problem Description
|
|
It was reported that the controller dc adapter did not make a smooth easy snap on connection.The controller dc adapter was replaced.No patient complications have been reported as a result of this event.
|
| |
|