| Model Number |
C0R37 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The event device is anticipated to return.A follow-up report will be provided upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Procedure performed: "laparoscopic gastric sleeve" event description: cer 1 of 2: (b)(4), cer 2 of 2: (b)(4)."happened in both first case and second case.Part of the seal fell off in the patient and they had to go retrieve it.Happened in the middle of the case.The surgeon was using the trocar." type of intervention: "procedure completed with new trocar" patient status: "they couldn't state anything about potential future complications, but nothing was wrong then.".
|
| |
|
Additional Manufacturer Narrative
|
|
The event unit was not returned to applied medical for evaluation.However, nine (9) sterile units were returned for evaluation.As the event unit was not returned, testing was unable to be performed and the complainant's experience could not be replicated or confirmed.The sterile units were visually inspected and no non-conformances were observed.Based on the description of the event, it is likely that the reported event was caused by non-axial insertion or removal of asymmetrical instruments through the trocar.Applied medical's instructions for use states that, "extra care should be used when inserting angular and asymmetrical instruments, such as 'j' hooks and clip appliers.All instruments should be centered axially when inserted through the seal to prevent tearing." applied medical will continue to monitor its vigilance system for trends and take appropriate actions, as necessary, to ensure the performance and safety of its products.This report is to follow up medwatch report #mw#5074110.
|
| |
|
Event or Problem Description
|
|
Additional information was received via email, from sales associate on tuesday 2-jan-2018: "materials said he sent back all items, so if they aren't in the box they must have disposed of them." additional information was received via mail, from fda medwatch #mw5074110 on 11jan2018: "applied medical kii optical access system: trocar was inserted into patient.Surgeon was using []clip appliers and seal of trocar broke off into patient's abdomen.Surgeon retrieved the seal.No adverse impact to the patient.Diagnosis or reason for use: surgery.Event abated after use stopped or dose reduced: yes.".
|
| |
|