Additional information, device evaluation the reported pipeline flex embolization device (ped) was received for analysis.As received, the ped was within the lumen of a microcatheter.No issue was encountered pushing the ped out of the microcatheter for further examination.Upon exiting the microcatheter, the ped braid was found fully opened.Damages, severe fraying at the distal segment and slight fraying at the proximal segment, were noted on the ped braid.No other anomalies were observed.Based on the findings, the report of ped braid "failure to open at the distal segment" could not be confirmed.The patient's moderate vessel tortuosity and damaged braid might have contributed to the reported issue.However, the actual cause of the reported clinical experience could not be determined.All products are 100% inspected for damage and irregularities during manufacture.No evidence was found to suggest that the devices failed to meet specifications the ped instruction for use states, "do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ if information is provided in the future, a supplemental report will be issued.
|