MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AESPIRE 7900; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AESPIRE 7900; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/28/2017
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the equipment and was not able to confirm the reported complaint.The bag/vent switch was replaced and the unit was returned to service.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported the unit lost the ability to mechanically ventilate.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AESPIRE 7900
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 
MDR Report Key7143834
MDR Text Key95669499
Report Number2112667-2017-02385
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K050626
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/28/2017
Initial Date FDA Received12/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/13/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-