MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. QUADRA ALLURE MP CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problem Failure to Interrogate (1332)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/24/2017
Event Type  malfunction  
Event Description
It was reported that it was not possible to interrogate a non-implanted device.The device was still inside the packaging and it was not possible to interrogate the device with a programmer that did not have the most recent software update.There was no patient involved in this event.
 
Manufacturer Narrative
No conclusion code available.The reported event of a device integration anomaly was not confirmed.The device was in normal mode when received for analysis.The error message seen in the field is consistent with programmer running older version of software, which does not meet the latest cyber security encryption requirements of the updated device firmware.Several analysis tests were performed including electrical, mechanical and environmental stress testing; all tests indicated the device exhibited normal characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ALLURE MP CRT-P
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7143852
MDR Text Key95729465
Report Number2017865-2017-36235
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2019
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000048755
Other Device ID Number05414734510097
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/24/2017
Initial Date FDA Received12/22/2017
Supplement Dates Manufacturer Received01/14/2018
Supplement Dates FDA Received02/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-