ABBOTT DIAGNOSTICS DIVISION CELL-DYN 1800 ANALYZER; AUTOMATED HEMATOLOGY ANALYZER Back to Search Results
Catalog Number 07H77-01
Medical Device Problem Code Incorrect Or Inadequate Test Results (2456)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.An evaluation is in process.
 
Event or Problem Description
The customer observed discrepant hemoglobin (hgb) results while using the cell-dyn 1800 analyzer.The following data was provided.Initial 7.4, repeat 8.6 g/dl no impact to patient management was reported.
 
Additional Manufacturer Narrative
Evaluation of the customer issue included a review of the complaint text, a search for similar complaints, an instrument service review, and labeling review.No adverse trend was identified for the customer issue.Service history review identified no contributing factors to the customer issue.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.Describe event or problem, update information.An evaluation is in process.
 
Event or Problem Description
The customer provided a previous hemoglobin result (10.0 g/dl) for the patient, however the specific test date was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL-DYN 1800 ANALYZER
Common Device Name
AUTOMATED HEMATOLOGY ANALYZER
Manufacturer (Section D)
ABBOTT DIAGNOSTICS DIVISION
4551 great america parkway
santa clara CA 95054
MDR Report Key7143853
Report Number2919069-2017-00170
Device Sequence Number939750
Product Code GKZ
UDI-Device Identifier00380740071141
UDI-Public00380740071141
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K030513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Medical Assistant
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number07H77-01
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 12/14/2017
Supplement Date Received by Manufacturer01/11/2018
02/20/2018
Initial Report FDA Received Date12/22/2017
Supplement Report FDA Received Date01/24/2018
02/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-