Catalog Number 0998-00-0800-53
Device Problems Charging Problem (2892); Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/02/2017
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this intra-aortic balloon pump (iabp) is not required to be reviewed per company standard operating procedure since the device manufacture date is greater than one year from the event date.A getinge field service engineer (fse) was dispatched and found that the batteries were not charging in slot 1 or 2.The fse replaced the defective batteries.The display mount would not stay down and the fse found that it was due to the mount screws being loose.The fse replaced the screws and verified that the mount retracts and locks in place.The fse verified the iabp calibrated and passed all functional and safety tests per factory specifications.The iabp was then returned to the customer cleared for clinical use.
 
Event Description
The customer reported that the display mount on the cardiosave intra-aortic balloon pump (iabp) does not stay down in locked position and the battery in slot 1 is not charging.This event occurred during service/test performed by the customer.No patient was involved and no adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7143968
MDR Text Key95927967
Report Number2249723-2017-01059
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-