ENCORE MEDICAL L.P. BIOLOX DELTA HIP; HEAD, FEMORAL, CERAMIC, BILOX DELTA, 44MM NEUTRAL Back to Search Results
Model Number 400-03-442
Health Effect - Clinical Code No Code Available (3191)
Date of Event 12/11/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient having bad bone quality.A 400-03-401 djo head was implanted in the revision and the shell and liner were replaced with another brand and not djo.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to poor bone quality.The previous surgery and the revision detailed in this investigation occurred 4.1 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to poor bone quality.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HIP
Common Device Name
HEAD, FEMORAL, CERAMIC, BILOX DELTA, 44MM NEUTRAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7145751
Report Number1644408-2017-01189
Device Sequence Number927183
Product Code LZO
UDI-Device Identifier00888912076203
UDI-Public(01)00888912076203
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K082844
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/27/2021
Device Model Number400-03-442
Device Catalogue Number400-03-442
Device Lot Number877B1065
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/11/2017
Supplement Date Received by Manufacturer01/30/2018
Initial Report FDA Received Date12/26/2017
Supplement Report FDA Received Date01/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
430-93-060, LOT 046H1011; 931-44-260, LOT 710F1017
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
-
-