| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional product code: dro.Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self-test for pacer and defib functions.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical canada; the malfunction was observed and the processor/bridge/pace board was replaced.The device was recertified and returned to the customer.The processor/bridge/pace board was returned to zoll medical us; the malfunction was duplicated and attributed to a fractured solder joint on a transformer on the processor/bridge/pace board.Analysis for reports of this type has not identified an increase in trend.
|
| |
|