Model Number ZXT225
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 and (b)(6) 2017.If explanted, give date: not applicable, as the lens remains implanted.However, there is a planned explant on (b)(6) 2018.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zxt225 20.0 diopter intraocular lens (iol) was implanted in the patient's right eye (od) on (b)(6) 2017.Reportedly, the patient is currently 3 months out post-operatively and there is a planned iol exchange to replace the lens with a toric iol for distance, (b)(6) 2018.Also, the patient is experiencing blurred vision and glare.Further information provided reports patient is making improvements.No additional information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the product testing could not be performed as the complaint device was not returned for evaluation.The reported complaint cannot be confirmed.Manufacturing record review: the manufacturing process record was evaluated and no deviation was found during process related to the complaint issue reported.There was no discrepancy found during the mrr (manufacturing record review).The product was manufactured and released according to the specifications.A search of complaints was performed on dec-12-2017.The search revealed that no similar complaints were received for this production order number.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to the manufacturer has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7146634
Report Number9614546-2017-01258
Device Sequence Number16375483
Product Code POE
UDI-Device Identifier05050474605794
UDI-Public(01)05050474605794(17)210705
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date07/05/2021
Device Model NumberZXT225
Device Catalogue NumberZXT225U200
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/28/2017
Supplement Date Received by Manufacturer01/15/2018
Initial Report FDA Received Date12/26/2017
Supplement Report FDA Received Date02/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-