| Model Number |
M0035431030 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/04/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil detached prematurely inside the microcatheter during manipulation.The coil was retrieved with a snare and the procedure was completed successfully.There were no reported clinical consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing date ¿ added.Device evaluated by mfg ¿updated.Summary attached - updated.Expiration date - added.Product available to stryker ¿ updated.Returned to manufacturer on ¿updated.The device history record review confirms that the device met all material, assembly and performance specifications.The subject device was returned with a snare device.The coil delivery wire was not returned for analysis.The main coil was returned snared within a third party device.Visual inspection of the device revealed that the main coil was kinked, stretched and prematurely detached/separated due to a physical break at the detachment zone.No other anomalies were noted.Functional testing could not be performed due to the damaged condition of the returned coil.Information available indicated that resistance was observed during advancement of the coil in the microcatheter.It is likely that the main coil was damaged during use thus limiting the performance of the coil during the clinical procedure.Therefore, an assignable cause of operational context has been assigned to this investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the coil detached prematurely inside the microcatheter during manipulation.The coil was retrieved with a snare and the procedure was completed successfully.There were no reported clinical consequences to the patient.
|
| |
|