| Model Number |
I7 |
| Medical Device Problem Codes |
No Display/Image (1183); Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
11/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.No additional evaluation has been performed as the issue was confirmed to be resolved by medtronic personnel through initial troubleshooting.Issue resolved via troubleshooting/.
|
| |
|
Event or Problem Description
|
|
A medtronic representative reported that, while outside of a procedure, the imaging system displayed that the touchscreen was searching for a dvi connection.The configuration of the fiber optic cable was manually adjusted to different ports on the navigation system to restore functionality.There was no patient present when this issue was identified.No additional information was provided.
|
| |
|