• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number I7
Medical Device Problem Codes No Display/Image (1183); Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No additional evaluation has been performed as the issue was confirmed to be resolved by medtronic personnel through initial troubleshooting.Issue resolved via troubleshooting/.
 
Event or Problem Description
A medtronic representative reported that, while outside of a procedure, the imaging system displayed that the touchscreen was searching for a dvi connection.The configuration of the fiber optic cable was manually adjusted to different ports on the navigation system to restore functionality.There was no patient present when this issue was identified.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM
Common Device Name
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
MDR Report Key7147472
Report Number1723170-2017-05302
Device Sequence Number1109754
Product Code HAW
UDI-Device Identifier00643169486645
UDI-Public00643169486645
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date (Section B) 12/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Model NumberI7
Device Catalogue Number9734060
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2017
Initial Report FDA Received Date12/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-