• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 11/13 S-ROM 40MM M SPEC+6; FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 11/13 S-ROM 40MM M SPEC+6; FEMORAL HEAD Back to Search Results
Catalog Number 136544500
Medical Device Problem Codes Material Disintegration (1177); Naturally Worn (2988); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Not Applicable (3189); No Code Available (3191)
Date of Event 12/04/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that it had a 40x58 metal on metal liner done in 2008.Elevated cobalt level was elevated at 5.3 mcg/l and shows osteolysis.Changed the metal line to poly and increased the head from +9 to +12.
 
Additional Manufacturer Narrative
Product complaint #
=
> pc-(b)(6).Investigation summary
=
> no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate. .
 
Additional Manufacturer Narrative
Product complaint #
=
> pc-(b)(4).
 
Event or Problem Description
Pinnacle litigation alleged friction and wear between the cobalt-chromium metal head and cobalt-chromium metal liner caused large amounts of toxic cobalt-chromium metal ions and particles to be released into plaintiffs blood.As a result he began experiencing severe pain, discomfort, inflammation, and injury.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
 
Event or Problem Description
Pfs alleges difficulty walking.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11/13 S-ROM 40MM M SPEC+6
Common Device Name
FEMORAL HEAD
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co
cork IN
EI 
MDR Report Key7147900
Report Number1818910-2017-52547
Device Sequence Number11126613
Product Code JDI
UDI-Device Identifier10603295033851
UDI-Public10603295033851
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K060031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2012
Device Catalogue Number136544500
Device Lot Number2437542
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2017
Supplement Date Received by Manufacturer01/25/2018
07/31/2018
02/14/2019
11/29/2019
Initial Report FDA Received Date12/27/2017
Supplement Report FDA Received Date01/25/2018
08/22/2018
02/27/2019
12/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient Weight87
-
-