• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/29/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative was unable to replicate the reported issue.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.
 
Event Description
A site representative reported that, while in a functional endoscopic sinus surgery (fess), the site was unable to register the patient for the procedure.It was reported that the navigation system displayed a message stating there the user did not maintain skin contact.It was reported that the image used in the procedure was taken on (b)(6) 2017.The surgeon opted to complete the procedure without the use of the navigation system.There was no reported delay to the procedure due to this issue.There was no known impact on patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7147924
MDR Text Key95837054
Report Number1723170-2017-05309
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age40 YR
Patient Weight75
-
-