• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX TI 5.0 ULTRABRAID; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72200755
Device Problem Material Rupture (1546)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/20/2017
Event Type  malfunction  
Event Description
It was reported that after implanted the anchor sutures with a twinfix, they were ruptured during knotting step.No patient injury was reported.
 
Manufacturer Narrative
Due to no product return, the complaint could not be confirmed.Definitive conclusions cannot be made without a device to evaluate.Accurate investigation and evaluation are not possible.The product met specifications upon release to distribution.No further actions pursued at this time.A dhr review was performed by querying the ncmr database for historical data.This search identified no ncmr¿s for this product number and lot.Complaint history search revealed no other complaints for this production lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX TI 5.0 ULTRABRAID
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key7148137
MDR Text Key95934221
Report Number1219602-2017-01596
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010577986
UDI-Public(01)03596010577986(17)220328(10)50655510
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K972326
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/28/2022
Device Model Number72200755
Device Catalogue Number72200755
Device Lot Number50655510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-