REVISION OPTICS RAINDROP NEAR VISION INLAY; CORNEAL INLAY Back to Search Results
Model Number 610-0001
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Loss of Vision (2139)
Date of Event 10/24/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The explanted inlay was returned to the manufacturer and subjected to visual microscopic inspection and dimensional analysis.The edge thickness and diameter were measured and found to be within specifications.Cuts and particles were observed on the device, but these findings are consistent with findings for corneal inlays that have been explanted since surgical instruments are required to remove the device from the eye and place it in a hydrated storage container for transport.The device history record review of the manufacturing lot was performed and there were no discrepancies or unusual findings related to the reported issue.Decreased visual acuity is listed in the device labeling as a known potential risk.(b)(4).
 
Event or Problem Description
The patient underwent uneventful implantation of the raindrop corneal inlay in the right eye on (b)(6) 2017.The inlay was explanted on (b)(6) 2017 in order to address decreased best corrected distance visual acuity (bcdva) from 20/15 (preoperatively) to 20/40+2.Post explant, bcdva improved to 20/30.The reason for decreased vision remains unknown.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Patient follow-up was requested from the surgeon, who provided the following additional information.The patient was last examined on (b)(6) 2017 and the best corrected distance visual acuity (bcdva) had improved to 20/20-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAINDROP NEAR VISION INLAY
Common Device Name
CORNEAL INLAY
Manufacturer (Section D)
REVISION OPTICS
25651 atlantic ocean dr.,
ste. a1
CA 92630 8835
MDR Report Key7148558
Report Number3005956347-2017-00167
Device Sequence Number1238135
Product Code LQE
UDI-Device Identifier10850394006013
UDI-Public(01)10850394006013
Combination Product (Y/N)N
Initial Reporter StateSD
Initial Reporter CountryUS
PMA/510(K) Number
P150034
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date09/16/2019
Device Model Number610-0001
Device Catalogue NumberRD1-1
Device Lot Number002999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/28/2017
Supplement Date Received by Manufacturer01/08/2018
Initial Report FDA Received Date12/27/2017
Supplement Report FDA Received Date01/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-