• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SPENCER ADVANIX¿ PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SPENCER ADVANIX¿ PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00537170
Medical Device Problem Code Kinked (1339)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.Device code relates to problem code for the reported event of "stent kinked".A visual evaluation of the returned device found that a section of the stent is bent and the pigtail is kinked.The stent also has several scratches in different locations.Based on the product analysis, most likely some procedural/anatomical factors were encountered during the procedure which may have limited the overall performance of the device.A device under tortuous conditions due to patient anatomy or customer use/manipulation/maneuvering can bend and damage the stent, also anatomical factors and the device condition could have resulted in issues of interaction with the guidewire and stent/delivery system.Therefore, ¿operational context¿ is selected as the most probable root cause for the complaint.A dhr (device history record) review was performed and no deviation was found.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
Note: this report pertains to one of three devices used during the same procedure.Refer to manufacturer report # 3005099803-2017-03865, 3005099803-2017-03869, 3005099803-2017-03866 for the associated device information.It was reported to boston scientific corporation that an advanix pancreatic stent was used during an endoscopic retrograde cholangio-pancreatography procedure performed on (b)(6) 2017.As the physician introduced a 7frx 11cm advanix pancreatic stent into the patient¿s pancreatic duct, he experienced some resistance.The stent was advanced into position and when the technician attempted to deploy the stent the delivery system and advanix pancreatic stent both buckled/crunched up which would not allow the stent to release from the catheter.The device was removed from the scope.They then attempted to place a 5frx9cm advanix pancreatic stent over a guidewire and push it into position with a hydratome.During the advancing of the 5frx9cm advanix pancreatic stent, it too buckled and had to be removed.Finally, a naviflex pancreatic delivery system was loaded with the 7frx9cm advanix pancreatic stent and attempted to be pushed into position.The stent was in position and the delivery system buckled during deployment.The physician had to use the elevator of the scope to pinch/hold the end of the stent in place during the deployment.The procedure was completed after placing the 7frx9cm advanix pancreatic stent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX¿ PANCREATIC STENT
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - SPENCER
780 brookside drive
spencer IN 47460
MDR Report Key7148572
Report Number3005099803-2017-03866
Device Sequence Number19109908
Product Code FGE
UDI-Device Identifier08714729856641
UDI-Public08714729856641
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K133700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2019
Device Model NumberM00537170
Device Catalogue Number3717
Device Lot Number20937981
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2017
Initial Report FDA Received Date12/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-