• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The representative reported that the poe injector for the navigation system was replaced.The hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The poe injector was returned to the manufacturer for analysis.The device was found to be fully functional with no problem found.The reported event could not be duplicated by medtronic personnel.
 
Event Description
A site representative reported that, while in a cranial resection, the navigation system displayed that the localizer was not connected after unplugging a microscope.Restarting the navigation system temporarily restored functionality, however the error reappeared.Further investigation found that the ethernet connections within the navigation system were loose and after tightening the system functioned as designed.There was no reported delay to the procedure due to this issue.There was no impact on patient outcome.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7148620
MDR Text Key95858103
Report Number1723170-2017-05324
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue Number9735665
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age80 YR
Patient Weight80
-
-