PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. ICT; SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED Back to Search Results
Model Number 728306
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: we have not completed our investigation of this event.We will file a follow-up emdr at the completion of the investigation.Internal cross reference: complaint (b)(4).
 
Event Description
The customer reported that the patient could not be heard from the gantry microphone.If the operator is unable to hear the patient due to a failed microphone, there is potential for injury to the patient.This issue has been determined to be a reportable event.This event is currently under investigation to determine the cause of the event.
 
Manufacturer Narrative
The customer reported that the patient could not be heard from the gantry microphone.The customer confirmed with the philips helpdesk that there was no patient impact and no harm as a result of this event.The philips field service determined that the microphone inside the breathing light assembly had failed and replaced the complete breathing light assembly to correct the issue.The system is fully functional and is in clinical use.Further investigation and evaluation of this issue determined this not to be a reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICT
Type of Device
SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
595 miner rd
cleveland OH 44143
Manufacturer Contact
derrick massuri
595 miner rd
cleveland, OH 44143
4404833000
MDR Report Key7149166
MDR Text Key95954785
Report Number1525965-2017-00159
Device Sequence Number1
Product Code JAK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K060937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number728306
Device Catalogue NumberNCTC780
Other Device ID NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/01/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-