Catalog Number DTDW-35-110-10
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/13/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The event is currently under investigation.A supplemental report will be provided upon conclusion.
 
Event or Problem Description
It was reported that during a procedure, the reuter disposable tip deflecting wire guide broke.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.  according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Additional information regarding event details, patient anatomy and outcome has been requested, but is not available at this time.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, documentation, manufacturing instructions, quality control, and trends of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REUTER DISPOSABLE TIP DEFLECTING WIRE GUIDE
Common Device Name
DQX WIRE, GUIDE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7149608
Report Number1820334-2017-04493
Device Sequence Number13925231
Product Code DQX
UDI-Device Identifier00827002269576
UDI-Public(01)00827002269576(17)221115(10)8270702
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberDTDW-35-110-10
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2017
Supplement Date Received by Manufacturer03/16/2018
Initial Report FDA Received Date12/27/2017
Supplement Report FDA Received Date04/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-