| Model Number |
ZCT600 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If explanted; give date: n/a (not applicable).The intraocular lens remains implanted.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the subject had the intraocular lens implanted on (b)(6) 2017, and a misalignment of 11 degrees from 93 degrees to 82 degrees was noted at the 1-month visit for the left eye.There was no patient complication and/or related injury.The outcome does not significantly interfere with activities of daily life.The patient is stable and doesn''t want to seek any treatment at this time.No additional information was provided to abbott medical optics.Note: patient had bilateral lens implants; therefore, two reports are being submitted, one for each impacted eye.This report captures the event for the left eye.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned to the manufacturing site as it remains implanted.Therefore, product testing could not be performed and the customer¿s reported complaint could not be verified.The manufacturing records for the product was reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.The directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
|
| |
|
Search Alerts/Recalls
|