• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SUMMIT BASIC CEMENTED SZ 4; SUMMIT HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SUMMIT BASIC CEMENTED SZ 4; SUMMIT HIP STEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 157006100
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 12/02/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address infection and loosening of the stem at the cement to implant interface.Competitor cement was used.He was brought to surgery for an i&d wash out and uni polar head exchange.During surgery, it was discovered that the cemented stem was loose and had a biofilm, mucus type film around it and needed to be removed.A new smaller stem and uni head were then implanted.Doi: (b)(6) 2017; dor: (b)(6) 2017.Unknown affected side.
 
Additional Manufacturer Narrative
(b)(4).No device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT BASIC CEMENTED SZ 4
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key7150596
Report Number1818910-2017-52593
Device Sequence Number13719619
Product Code JDI
UDI-Device Identifier10603295059776
UDI-Public10603295059776
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
K023453
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 12/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number157006100
Device Lot NumberD17022203
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/02/2017
Supplement Date Received by Manufacturer07/30/2018
Initial Report FDA Received Date12/28/2017
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age79 YR