ENCORE MEDICAL L.P. EXPRT KNEE; INSERT, REVISION EXPRT, SZ A X 17MM, E-PLUS Back to Search Results
Model Number 315-0A-717
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/13/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to the patient rupturing their extensor mechanism.The surgeon swapped the patient's poly out for a 19.5 mm during the procedure.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to the patient's ruptured extensor mechanism.The previous surgery and the revision detailed in this investigation occurred 3.5 months apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.Initial or prolonged hospitalization was required.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.The device was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was due to patient's ruptured extensor mechanism.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the event.There are multiple factors that may contribute to the event that are outside the control of djo surgical such as: an unbalanced joint, inadequate soft tissue support, excessive load, patient activities or trauma.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT KNEE
Common Device Name
INSERT, REVISION EXPRT, SZ A X 17MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7151005
Report Number1644408-2017-01230
Device Sequence Number140687
Product Code JWH
UDI-Device Identifier00888912113922
UDI-Public(01)00888912113922
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/10/2022
Device Model Number315-0A-717
Device Catalogue Number315-0A-717
Device Lot Number005R1013
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/13/2017
Supplement Date Received by Manufacturer02/13/2018
Initial Report FDA Received Date12/28/2017
Supplement Report FDA Received Date02/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
-
-