| Catalog Number |
93900 |
| Medical Device Problem Codes |
Crack (1135); Use of Device Problem (1670); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/01/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Device labeling: 5.Precautions ¿ juvéderm® ultra plus xc is packaged for single-patient use.Do not resterilize.Do not use if package is opened or damaged.¿ failure to comply with the needle attachment instructions could result in needle disengagement and/or product leakage at the luer-lok® and needle hub connection.8.Instructions for use b.Health care professional instructions 8.If the needle is blocked, do not increase the pressure on the plunger rod.Instead, stop the injection and replace the needle.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported the event of during patient injection with one syringe of juvéderm® ultra plus xc the needle disengaged, the luer lock cracked, and product spilled out at the luer lock.Patient contact was made.No injury to the patient or injector was reported.The needle from the package was used.This was the initial use of the syringe.
|
| |
|