• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Imprecision (1307)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2017
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to site to test the equipment. Representative was unable to replicate the reported issue. A system checkout was performed and the hardware, software, and instruments passed the system checkout. The system was found to be fully functional.
 
Event Description
A medtronic representative reported that during cranial resection, the surgeon stated navigation was off and slightly deep. The procedure was completed with the use of navigation. There was no delay. No impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S7 SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7151409
MDR Text Key96010635
Report Number1723170-2017-05343
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-