• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 35 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number PTA5-35-80-4-4.0
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This event is currently under investigation.A supplemental report will be provided upon conclusion.
 
Event or Problem Description
It was reported that during a procedure, the advance 35 lp low profile balloon catheter ruptured longitudinally at the proximal end of the balloon.The inflation pressure used was 12 atmosphere (atm); the balloon ruptured in the patient's fistula.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.  according to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Corrected information: narrative: the balloon ruptured at the anastomosis in the patient's fistula.Investigation - evaluation: a review of the complaint history, drawings, device history record, documentation, instructions for use (ifu), manufacturing instructions, specifications, and quality control was conducted during the investigation.The complaint device was not returned; therefore no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Numerous design verification and validation activities have been performed to ensure that this device meets design requirements.Review of the device history record of the finished product shows one nonconforming event that could contribute to this failure mode; the associated device was rejected and scrapped.A complaint history search revealed that there were no other reported complaints for this lot number.Per the ifu," the balloon is manufactured from an extra-thinwall, high-strength, minimally compliant material.Particular care should be taken in handling the balloon to prevent damage.Do not exceed rated burst pressure.Rupture of balloon may occur.Over-inflation may cause rupture of the balloon, with resultant damage to the vessel wall.The burst pressure data was analyzed using factors for a one-sided tolerance to determine with a 95% confidence that 99.9% of these balloons would not burst at or below the calculated rated burst pressure.In heavily scarred access site, use of an introducer sheath is recommended.Use only the recommended balloon inflation medium.Choose a balloon appropriate to lesion length and vessel diameter.Upon removal from package, inspect the product to ensure no damage has occurred." based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time; however, appropriate measures have been initiated to address this failure mode.Per the quality engineering risk assessment, no further action is required.Appropriate personnel have been notified and monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 35 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7151586
Report Number1820334-2017-04573
Device Sequence Number16178104
Product Code DQY
UDI-Device Identifier10827002523323
UDI-Public(01)10827002523323(17)200927(10)8249391
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K091527
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTA5-35-80-4-4.0
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer02/16/2018
Initial Report FDA Received Date12/28/2017
Supplement Report FDA Received Date03/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-