| Catalog Number |
963341 |
| Medical Device Problem Code |
Bent (1059)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
04/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the silver peg on the bottom of the tc3 rp insert is crooked/bent.Therefore it no longer sits precisely on the top of the tibial tray trial or real mbt rev tibial tray.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination of the returned device confirmed the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|