The customer reported via phone call that the customer had high blood glucose level.The customer¿s blood glucose level was 300 mg/dl at the time of incident and current blood glucose level was 57 mg/dl.The customer was treated with manual injection for high blood glucose level.The customer¿s blood glucose level was around 300 mg/dl and 500 mg/dl, in the morning it was 599 mg/dl and went down to 96 mg/dl after injecting manual injection.The customer reported that she did not have any symptom related to high blood glucose level however, had stomach virus.The customer reported that the insulin pump was under delivering as the blood glucose level was not going down even after bolusing multiple times.The drive support cap was normal.The customer removed reservoir, rewound, reinserted reservoir and ran manual prime and filled tubing with current set and insulin exited.Customer was advised to discontinue use of the device and revert to a back-up plan.The customer was advised that the device would be replaced and agreed to return the insulin pump for analysis.
|
Device received with intermittent button response due to flattened (escape and act ) button dome switch (no crease) and partial unlocked lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test , rewind, basic occlusion, occlusion, prime or compromised force sensor system alarm and excessive no delivery test or verify under delivery anomaly due to buttons not responding.Device received with cracked battery tube threads, cracked lcd window, minor scratched lcd window, cracked reservoir tube lip, cracked case display window corner, cracked belt clip slot and cracked case reservoir tube window corner.
|