Model Number NB5STF
Device Problems Device Packaging Compromised (2916); Packaging Problem (3007)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/09/2017
Event Type  malfunction  
Event Description
According to the reporter, pre-operatively laparoscopic radical nephrectomy, while preparing to open the trocar for puncturing, the seal was damaged on the product packaging.They changed the surgical instruments to complete the case.The patient had no injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted; trocar, cannula and circular seal appeared intact.The obturator was received inserted into the cannula; prolonged insertion can cause the envelope seal to become stretched.Pmv performed functional testing: the port failed an air leak test.The obturator was inserted and removed from the cannula cleanly.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPORT
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7152420
MDR Text Key96008833
Report Number9612501-2017-06848
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier10884521076747
UDI-Public10884521076747
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K100548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberNB5STF
Device Catalogue NumberNB5STF
Device Lot NumberJ6G1892X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2017
Initial Date FDA Received12/28/2017
Supplement Dates Manufacturer Received03/03/2018
Supplement Dates FDA Received03/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-