(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The reported patient effects of angina and thrombosis are listed in the xience alpine everolimus eluting coronary stent systems instructions for use as known patient effects of coronary stenting procedures.A conclusive cause for the reported patient effects and the relationship to the product, if any, cannot be determined and there is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
It was reported that during balloon pre-dilatation of the 90% stenosis in the heavily calcified, distal left circumflex coronary artery, the patient experienced st elevation.A 2.50x38 mm xience alpine stent was successfully deployed, reducing stenosis to 0%.Post-dilatation was not performed.About twenty minutes post procedure, the patient experienced chest pain with st elevation, and was brought back to the cath lab.Activated clotting time (act) was noted to be at 214 seconds with a timi 3 blood flow.Additional heparin was administered.An occlusion/clot was unable to be visualized; however, the physician suspected either a clot or unseen calcium behind the stent may have caused the st elevation.Post stent balloon dilatation was performed and the chest pain and st elevations resolved.An intravenous infusion of integrilin was started.In the opinion of the physician, the stent did not cause or contribute to the patient event.The physician believes the event was due to inadequate anti-coagulation and residual calcium in the vessel.The patient was discharged home on (b)(6) 2017 and was doing great.There was no additional information provided.
|