• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PROCLAIM 5 ELITE; SCS IPG Back to Search Results
Model Number 3660
Device Problem Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The device is included in the neuromodulation implantable pulse generator (ipg) inaccurate elective replacement indicator advisory notice issued by abbott on (b)(6) 2017.The manufacturer has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported (b)(6) that diagnostics indicated the elective replacement indicator (eri) triggered earlier than intended.The device has the appropriate level of battery voltage to provide therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 5 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key7153455
MDR Text Key96019335
Report Number1627487-2017-08756
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
P001032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2018
Device Model Number3660
Device Lot Number5573674
Other Device ID Number05415067020192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2017
Initial Date FDA Received12/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487/09/12/2017/001-C
Patient Sequence Number1
-
-