| Model Number |
TRUBALANCE |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/08/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 191 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 150 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was performed by the customer fasting and produced test results of 100 mg/dl and 85 mg/dl using truebalance meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 04/14/2019 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called in states the meter is reading high.Customer states the meter read 191 mg/dl fasting.
|
| |
|