| Medical Device Problem Code |
Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and evaluated by product analysis on 08-dec-2017 with the following findings: a visual inspection of the pump showed a crack in the battery compartment from the threads to the case seal, left of the grip pad.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2017, the reporter contacted animas, alleging a casing/condition (cracked/damaged casing) issue.It was reported that the battery compartment was damaged.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
| |
|
Search Alerts/Recalls
|