It is unknown when the event occurred.(b)(4).This report is for helical blade/unknown lot.Lot number are unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
It was reported that patient had a revision on (b)(6) 2017, due to helical blade was sticking out.Patient was implanted with trochanteric fixation nail, date unknown.Patient began to have pain of the hip at the joint, x-ray was taken, date unknown, it was noticed that the helical blade was sticking out significantly lateral because the bone had collapsed.It was reported that the device functioned correctly by sliding and not cutting out the femoral head, but caused the patient pain on the lateral side of the hip.Helical blade was removed and a shorter was implanted.Concomitant device(s) reported: tfn nail (part # unknown, lot # unknown, quantity 1).This complaint involves one (1) device.This report is for 1 (one) helical blade.(b)(4).
|