| Model Number |
G48428 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is available for evaluation but has not yet been received by the manufacturer.(b)(4).The event is currently under investigation.A follow up report will be submitted upon receipt of additional information or completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that upon opening the zenith flex aaa endovascular graft bifurcated main body in the sterile field, the physician noticed what appears to be a hair stuck in the peel away.A new device was used to complete the procedure.There were no adverse effects to the patient reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation/evaluation: the actual complaint device was returned and a visual inspection was performed.A review of the device history record, complaint history, and quality control data was also conducted.Reviewing the images of the complaint device revealed the presence of a dark fiber contained within the peel away sheath.The device was returned for analysis.The visual inspection of the device found no foreign matter on the device.It is likely that the fiber detached during shipping and or handling back to the manufacturer.The device history record was reviewed and no non-conformances were noted.A review of complaint history records revealed this complaint to be the only reported complaint associated with the complaint lot number 7776100.Each device is shipped with instruction for use (ifu) listing the indications for use, contraindications, warnings and precautions.It is feasible to suggest the foreign matter was introduced during a manufacturing process.Per the quality engineering risk assessment, no further action is required.The appropriate cook personnel have been notified of this event.Monitoring will continue to be performed for similar complaints.
|
| |
|
Search Alerts/Recalls
|