| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report # (b)(4).Returned meter evaluated with no defect found.Test strip vial returned empty - unable to test.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 303 mg/dl.The customer's expected fasting blood glucose test result range is 100 - 130 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was performed by the customer fasting and produced test results of 135 mg/dl and 161 mg/dl using truemetrix air meter.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 11/17/2018 and open vial date is 10/01/2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called in states the meter is reading high.
|
| |
|