• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTERLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC INTERLINK BLOOD RECIPIENT SET; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number 2C6750H
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/01/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported an air in line alarm occurred on a sigma spectrum pump while using an interlink blood recipient set.This event occurred during patient infusion.Blood was dripping freely, but when placed on the iv pump, it would constantly beep and say ¿air in line¿.There was no visible issue with the tubing that they could see, there was no air in line, the pump appeared to be clean and in good working order.The issue was resolved after they had changed the tubing.The pump would beep with the air in line alarm within a minute of starting the infusion.All connections were checked to ensure they were secure prior to infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not returned; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK BLOOD RECIPIENT SET
Common Device Name
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
MDR Report Key7155431
Report Number1416980-2017-10074
Device Sequence Number19327661
Product Code BRZ
UDI-Device Identifier00085412073583
UDI-Public(01)00085412073583
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K993120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/23/2022
Device Catalogue Number2C6750H
Device Lot NumberDR17I22014
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/05/2017
Supplement Date Received by Manufacturer01/11/2018
Initial Report FDA Received Date12/29/2017
Supplement Report FDA Received Date01/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIGMA SPECTRUM PUMP
-
-