MAUDE Adverse Event Report: COOK INC MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET; DYB INTRODUCER, CATHETER Back to Search Results
Catalog Number MPIS-501-10.0-SC-SST
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).This event is currently under investigation.A supplemental report will be provided upon conclusion.
 
Event or Problem Description
It has been reported that during an unspecified procedure, a portion of the micropuncture transitionless stiffened cannula access set device separated inside the patient.The separated portion was successfully retrieved with a snare.No unintended section of the device remain inside the patient¿s body.According to the initial reporter, there have been no adverse effects to the patient due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the complaint history, device history record, and quality control of the device was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed.However, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPUNCTURE TRANSITIONLESS STIFFENED CANNULA ACCESS SET
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7155454
Report Number1820334-2017-04584
Device Sequence Number73051
Product Code DYB
UDI-Device Identifier00827002480063
UDI-Public(01)00827002480063(17)201026(10)8327376
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberMPIS-501-10.0-SC-SST
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/27/2017
Supplement Date Received by Manufacturer04/06/2018
Initial Report FDA Received Date12/29/2017
Supplement Report FDA Received Date04/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-