Model Number MMT-751NAB
Device Problems Device Alarm System (1012); No Display/Image (1183); Low Battery (2584)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they had blank display.The customer¿s current blood glucose level was 390 mg/dl at time of incident.The customer stated that the insulin pump was getting low battery and the insulin pump stopped working.They replaced the battery, but also had a blank screen.The troubleshooting was performed and was advised to insert a new battery.Customer stated that the display did not return.The customer was advised to discontinue use of the pump and revert to backup plan per health care professional.The customer was advised that pump will need to be replaced.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Pump received with blank display, problem isolated to interface board.Unable to confirm low battery alert and unable to perform any tests due to blank display.Pump received with cracked reservoir tube lip, cracked case near display window corners, cracked on display window and minor scratches on display window noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAB
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7155924
MDR Text Key96124400
Report Number2032227-2017-78566
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169503670
UDI-Public(01)00643169503670
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 05/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAB
Device Catalogue NumberMMT-751NAB
Device Lot NumberA4751NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received05/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age29 YR
Patient Weight180
-
-