Model Number 15 |
Device Problem
Failure to Power Up (1476)
|
Patient Problem
No Information (3190)
|
Event Date 11/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device was not returned to physio-control for evaluation.The device will be sent to a third-party service agent.Physio continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
|
Event Description
|
It was reported to physio-control that the customer's device would not power on.The device was powered with ac, but wouldn't charge its batteries.The battery led was off and the auxiliary led flashed.After all the cables were disconnected and reconnected, the device started charging the batteries again.It was not reported if there was any patient use associated with the reported event.
|
|
Manufacturer Narrative
|
The device was eventually returned to physio-control.Physio evaluated the device but could not reproduce the reported issue.The device was extensively tested without observing any discrepancies.As soon as proper device operation has been confirmed through functional and performance testing, the device will be returned to the customer.A cause of the reported issue could not be confirmed.
|
|