Medical Device Problem Code Protective Measures Problem (3015)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2017, the reporter contacted animas, alleging a history/settings (suspend) issue.This complaint is being reported because suspending or resuming of the pump without an alarm may result in over or under delivery.
 
Additional Manufacturer Narrative
Follow-up #1: date of submission: 26-jan-2018.Device evaluation: the device has been returned and evaluated by product analysis on 04-jan-2018 with the following findings: a review of the pump¿s black box and alarm history showed evidence of multiple manual suspend alerts.During testing, the pump successfully completed the rewind, load, and prime steps without any interruptions or issues.The pump successfully completed the 24 hour duration test without interruptions or issues.The pump was manually suspended, resumed correctly, and adequately recorded the suspend event in the pump history.The original complaint of a pump suspend issue was unable to be duplicated.There was no defect found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7156749
Report Number2531779-2018-00009
Device Sequence Number1213934
Product Code OYC
UDI-Device Identifier10840406100099
UDI-Public10840406100099
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 12/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2017
Was the Report Sent to FDA? Yes
Device Age23 MO
Initial Date Received by Manufacturer 12/05/2017
Supplement Date Received by Manufacturer12/05/2017
Initial Report FDA Received Date01/02/2018
Supplement Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight150
-
-