• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 8MM TI STRAIGHT RADIAL STEM 28MM-STERILE; PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 8MM TI STRAIGHT RADIAL STEM 28MM-STERILE; PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER Back to Search Results
Model Number 04.402.008S
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Failure of Implant (1924); Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional narrative: patient¿s weight is unknown.Initially implanted earlier this year in 2017.Complainant device is not expected to be returned for manufacturer review/investigation.A device history record (dhr) review was performed: part #: 04.402.008s, lot#: 7607062 (sterile) - 8mm ti straight radial stem 28mm-sterile, (b)(4), manufacturing date: 28-jul-2014, expiration date: 30-jun-2019, components: part 21014, lot: 6190711.Raw material lot number 7557656 reviewed.(b)(4).(b)(4) product certification for titanium meets specification.Inspection sheet for incoming final inspection meets specification.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Sterility documentation was reviewed and determined to be conforming.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, this complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that a synthes radial head implant was removed on (b)(6) 2017 after fixation was nearing failure due to osteolysis around the stem as evidenced by lucency visible on the pre-op x-rays.The radial head and stem were removed intact.The patient was not implanted with any new hardware.No surgical delay.The implants were initially implanted earlier this year in 2017.Concomitant devices reported: 22mm cocr radial head standard height/12.5mm-sterile (part # 09.402.022s, lot # 7068118, quantity # 1).This report is for one (1) 8mm ti straight radial stem 28mm-sterile.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8MM TI STRAIGHT RADIAL STEM 28MM-STERILE
Common Device Name
PROSTHESIS, ELBOW, HEMI-RADIAL, POLYMER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7156765
Report Number2939274-2018-50003
Device Sequence Number10962879
Product Code KWI
UDI-Device Identifier10886982128058
UDI-Public(01)10886982128058(17)190630(10)7607062
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K112030
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Model Number04.402.008S
Device Catalogue Number04.402.008S
Device Lot Number7607062
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2017
Initial Report FDA Received Date01/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1124-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-