• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH, INC. TRUEMETRIX; BLOOD GLUCOSE SYSTEM Back to Search Results
Model Number TRUEMETRIX
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/07/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Most likely underlying root cause: (b)(4)-user's test strip had poor storage.(kitchen) test strip udi# (b)(4).
 
Event or Problem Description
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 317 and 401 mg/dl.Customer is also concerned with erratic blood glucose test results from back to back blood tests of 187 and 221 mg/dl.The customer's expected fasting blood glucose test result range is 150 - 220 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2017, a back to back blood test was not performed by the customer.The product is not stored according to specification and is stored in the kitchen.The test strip lot manufacturer's expiration date is 11/17/2018 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history (time not set; date was not set correctly for earlier results): result 1 :194mg/dl date: (b)(6) 2017 time:09:40pm fasting.Result 3 :187mg/dl date: (b)(6) /2017 time:12:11pm fasting.Result 4 :221mg/dl date: (b)(6) 2017 time:12:08pm fasting.Result 2 :317mg/dl date: (b)(6) 2017 time:09:43pm fasting.Result 2 :401mg/dl date: (b)(6) /2017 time:10:34pm fasting.Customer is primarily concerned with the 317mg/dl and 401 mg/dl readings.Customers time is not set correctly on the meter.Customers date was not set correctly for earlier readings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7156771
Report Number1000113657-2018-00002
Device Sequence Number101020
Product Code NBW
UDI-Device Identifier00093351018992
UDI-Public(01)00093351018992
Combination Product (Y/N)N
Initial Reporter StateOH
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 01/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/17/2018
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMU2476
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2017
Initial Report FDA Received Date01/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-