| Model Number |
TRUETRACK |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/07/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Product not yet returned for evaluation.Most likely underlying root cause: mlc-58-user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with test results from results obtained of 66, 72, 58 (all fasting) and 86 (non-fasting) mg/dl.The customer's expected fasting blood glucose test result range is 120 - 125 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The customer declined to perform a back to back blood test during the call on (b)(6) 2017.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 04/30/2019 and open vial date is (b)(6) 2017.Customer's test strips are not impacted by recall.The meter memory was reviewed for previous test result history (date/time not set; results from log book): (b)(6).
|
| |
|